Kimberly-Clark/Ballard medical R2 triangle shaped non-radiolucent multifunction electrode products for adults.

FDA Recall Z-1242-06 · Posted 2006-07-28

Recall StatusTerminated
Product CodeMKJ
Date Initiated2006-05-18
Date Terminated2007-05-14
Product Quantity95,639 electrode sets

Reason for Recall

The tinfoil in the triangle electrodes has the potential to develop cracks which may hinder electrical flow and product performance during defibrillation.

Firm Action

Consignees were first contacted by phone with follow up letter sent via fax or mail on May 18, 2006. International consignees will be notified through the firm's regional offices.

Root Cause

Device Design

Distribution Pattern

Worldwide ( including USA, Switzerland, Germany, UK, Canada, Australia, Netherlands, Italy, Greece, Spain, Hong Kong, Hungary, Denmark, Singapore, New Zealand, Ireland, Korea)