HealthPax Cranial Electrotherapy Stimulator

FDA Recall Z-1242-2007 · Posted 2007-09-26

Recall StatusTerminated
Product CodeJXK
Date Initiated2007-07-19
Date Terminated2008-06-03
Product Quantity212 units

Reason for Recall

Peak voltage test result not in acceptable limits.

Firm Action

The recalling firm sent a recall letter 07/30/2007 to all their customers. The letter indicated that it is necessary to return the product for testing and the device will either be returned or if necessary replaced. Each unit will be tested with an oscilloscope device to determine whether it meets or fails design specifications. The recalling firm will reimburse shipping costs.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Canada, Ireland, and India.