Terminated

Product is SoftPath ASCII Software Releases 2.3.0 and 4.3.7. The software is used in pathology labs for administrative and clinicial documentation and data processing.

FDA Recall Z-1243-06 · Posted 2006-07-27

Recalling FirmSCC Soft Computer
Recall StatusTerminated
Product CodeJQP
Date Initiated2004-10-27
Date Terminated2010-07-19
Product Quantity4

Reason for Recall

Text from one case is overwriting another. This could cause an incorrect diagnostic report to be sent to a clinician/surgeon.

Firm Action

Consignees were notified and the products were corrected by the firm.

Root Cause

Other

Distribution Pattern

Product was distributed to 4 clients, two in the United States and two in Canada.