Focalyx Fusion

FDA Recall Z-1243-2026 · Posted 2026-02-03

Recall StatusOpen, Classified
Product CodeLLZ
Date Initiated2025-12-23
Product Quantity18

Reason for Recall

Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.

Firm Action

On 12/23/2025, correction notices started being emailed to customers informing them of the following: You will be contacted to schedule installation of the new version. An IFU will be provided at time of installation. Complete and return the response form via email to [email protected] If there are any questions, contact the firm at: focalyxtechnologies.com/contact, [email protected], (888) 636 2259, (Sel. 3 Technical Support)

Root Cause

Software change control

Distribution Pattern

Worldwide - US Nationwide distribution in the states of NY, MA, FL and the countries of ES, VE, DO.