CA 15-3 Controls, list 9C08-10; an In Vitro diagnostic for use with the AxSYM or IMx CA 15-3 reagents, calibrators and instruments; each kit contains 2 bottles with 115D8:DF3 reactive determinants (human) prepared in TRIS buffer with protein (bovine) stabilizers to yield the following assay values (U/ml): Control L - 35 and Control H - 150; Abbott Laboratories, Abbott Park, IL 60064

FDA Recall Z-1244-06 · Posted 2006-07-18

Recall StatusTerminated
Product CodeMOI
Date Initiated2006-06-09
Date Terminated2008-01-18
Product Quantity1794 kits

Reason for Recall

The CA 15-3 Controls replicates out of range high resulting in invalid runs.

Firm Action

Abbott initiated a field correction recall by letter dated 6/9/06 and sent on 6/12/06 to all customers receiving the affected controls. The accounts were informed of the increase in invalid runs due to controls outside of the package insert range with this lot of CA 15-3 Controls. The accounts were requested to assess the inventory for the affected lot of controls, discontinue use of the lot and destroy the remaining inventory of the lot, recording the number of kits on the customer reply form and faxing it to Abbott by 6/21/06, indicating that the letter was received and understood. Credit will be issued for the number of kits destroyed, as indicated on the reply sheet. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688.

Root Cause

Other

Distribution Pattern

Worldwide, including U.S.A., Puerto Rico, Canada, Mexico, Ecuador, Chile, Argentina, Brazil, Germany, Japan, SIngapore, Hong Kong, Thailand, South Korea and Australia.