Coulter LH500 Hematology Analyzer: Part Numbers 178832, 178833, 178834
FDA Recall Z-1245-06 · Posted 2006-07-18
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeLOQ
Date Initiated2006-04-17
Date Terminated2012-04-19
Product Quantity759
Reason for Recall
Beckman has confirmed that erroneous results could be reported when the workstation database crashes.
Firm Action
A Product Corrective Action (PCA) letter was sent week of April 17, 2006 via US mail. The letter outlines action to be taken to avoid potential erroneous results. Firm targeting 75% effectivenss checks.
Root Cause
Other
Distribution Pattern
Nationwide and Canada