Zimmer TM-400 Broach Handle; Part No. 96-210-10001. The product is a surgical instrument for use in conjunction with Zimmer TM-400 Trabecular Metal implants during spinal surgery. The Broach Handle threads onto various broaches and trial implants. The handle with these broaches or trials attached is used to prepare and measure a vertebral space to accept a TM-400 Trabecular Metal implant.
FDA Recall Z-1246-2009 · Posted 2009-07-07
Advertisement
Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeHTQ
Date Initiated2009-04-29
Date Terminated2010-10-08
Product Quantity435
Reason for Recall
The broach handle may fracture at the threaded tip during use, which would result in surgical intervention to retrieve the broach or the trial as the threaded tip fragment prevents attachment of another handle to retrieve the device.
Firm Action
Distributors and user accounts were notified by "Urgent: Medical Device Recall" letter dated April 29, 2009. The letter described the affected product, reason for recall, health risks, and action for consignees.
Distributors were instructed to retrieve the devices and to return them to Zimmer no later than May 13, 2009. This instrument recall will result in an inability to implant the associated devices until a new instrument becomes available.
For questions or assistance about the recall please contact Zimmer Spine by calling 800-777-7505.
Root Cause
Device Design
Distribution Pattern
Nationwide, Austria, Australia, Belgium, Canada, Finland, Germany, Greece, Korea, South Africa, Sweden, Switzerland, Taiwan and the United Kingdom.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.