Hamilton brand RAPHAEL Ventilator (Software Versio 2.2x), Product is distributed by Hamilton Medical, Inc., 4990 Energy Way,Reno, NV 89502-Serial Numbers 2975 to 5360,
FDA Recall Z-1249-06 · Posted 2006-07-22
Advertisement
Recalling FirmHamilton Medical Inc
Recall StatusTerminated
Product CodeCBK
Date Initiated2006-05-24
Date Terminated2006-12-21
Product Quantity2,385 for all models
Reason for Recall
ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
Firm Action
On May 24, 2006 the manufacturer initiated this recall. The Recalling establishment/US Initial Importer/Distributor-received the manufacturer''s notice on May 31, 2006 and initiated their recall on June 1, 2006. The US distributor has issued a Medical Device Field Correction.
Root Cause
Other
Distribution Pattern
Worldwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.