Terminated

Synthes InSitu Bender Cutter Kit Narrow Sterile Catalog number 530.522S, used with Synthes Rapid Resorbable Fixation System for craniofacial skeleton fracture repair and reconstructive procedures.

FDA Recall Z-1249-2009 · Posted 2009-06-01

Recalling FirmSynthes Spine
Recall StatusTerminated
Product CodeHRS
Date Initiated2008-12-18
Date Terminated2009-12-15
Product Quantity40 units

Reason for Recall

The bender is heating prior to activation of the power button, requiring disassembly to stop heating.

Firm Action

All affected Synthes CMF Sales Consultants were notified of the Medical Device Recall by phone on 12/19/08 and by letter sent via return receipt email on 12/22/08. All affected hospitals were notified by letter mailed on 12/22/08 via USPS certified mail.

Root Cause

Process control

Distribution Pattern

Nationwide distribution: AL, CA, KS, LA, MD, NJ, NV, NY, OR, PA, TX, and UT