Hamilton brand RAPHAEL Silver Ventilator (Software Version 2.2xS) Product is distributed by Hamilton Medical, Inc., 4990 Energy Way, Reno, NV 89502

FDA Recall Z-1250-06 · Posted 2006-07-22

Recalling FirmHamilton Medical Inc
Recall StatusTerminated
Product CodeCBK
Date Initiated2006-05-24
Date Terminated2006-12-21
Product Quantity2,385 for all models

Reason for Recall

ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.

Firm Action

On May 24, 2006 the manufacturer initiated this recall. The Recalling establishment/US Initial Importer/Distributor-received the manufacturer''s notice on May 31, 2006 and initiated their recall on June 1, 2006. The US distributor has issued a Medical Device Field Correction.

Root Cause

Other