Elekta Leksell Stereotactic System, used for localization (spatial reference) for cranial surgery using X-ray or CT and MRI Image data.
FDA Recall Z-1251-2011 · Posted 2011-02-15
Advertisement
Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeHAW
Date Initiated2009-10-02
Date Terminated2011-02-17
Product Quantity379 units
Reason for Recall
New protocols for MR sequences may result in higher RF energies deposited during MR scanning, generating heat in uninsulated fixation posts.
Firm Action
Elekta sent Important Notice Affected Product: Fixation Posts, A331, dated June 22, 2009.
The Important Notice will be added to the Customer User manual upon receipt.
Root Cause
Labeling design
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.