In Vitro Diagnostic (IVD) Panel BioFire BCID2 Panel - REF RFIT-ASY-0147
FDA Recall Z-1251-2023 · Posted 2023-03-15
Advertisement
Recalling FirmBioFire Diagnostics, LLC
Recall StatusOpen, Classified
Product CodePEN
Date Initiated2023-02-01
Product Quantity2 kits
Reason for Recall
Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.
Firm Action
On 02/01/2023, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via email to customers informing them of a "Temperature Out of Range" issue with multiple products, as a result from Hurricane Ian (10/2022)
Customers are Instructed to:
Immediately examine their inventory for product identified in this recall.
1. Discontinue use and discard any remaining product in their possession that is subject to this recall. bioMerieux will replace the product at no charge in accordance with BIOFIRE s standard limited warranty.
2. Complete the accompanying Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to bioMerieux so that bioMerieux may acknowledge their receipt of this notification.
3. If customers have further distributed this product, please identify any recipients, and notify them at once of this product recall.
For questions or concerns, contact customer support department at [email protected] or via telephone by dialing +1.800.736.6354 and selecting option 5 for Product Technical Support
Root Cause
Environmental control
Distribution Pattern
U.S. Nationwide distribution in the states of FL and GA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.