Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

CAROLINA(R) LIQUID CHEMISTRIES CORP. ETHA RGT. KIT, IVD, REF ST321

FDA Recall Z-1252-2019 · Posted None

Recall StatusTerminated
Product CodeDIC
Date Initiated2019-01-16
Date Terminated2021-02-02
Product Quantity14 kits

Firm Action

The firm initiated the recall by email on 01/16/2019. The notice requested that the consignee discard the product.

Root Cause

Nonconforming Material/Component

Distribution Pattern

NY. NH, ME, NC, TX, AR, SC, CT, IL, CA, NJ

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑