Qwix 4.3mm Stabilization Screws - 32mm

FDA Recall Z-1253-2010 · Posted 2010-04-07

Recall StatusTerminated
Product CodeHWC
Date Initiated2010-02-08
Date Terminated2011-10-27

Reason for Recall

The screws were etched incorrectly as 34mm instead of the correct 32mm.

Firm Action

Consignees were notified via conference calls and/or FedEx mailing. Recall aknowledgement forms were provided with written notification. This requires consignees to review their inventory and to indicate any affected lots . This form requires a signature and must be returned.

Root Cause

Labeling mix-ups

Distribution Pattern

Nationwide distribution - Alabama, Arizona, Colorado, Connecticut, Maryland, Michigan, New England, New Jersey, North Carolina, Ohio, South Carolina, Nebraska, and Utah.