G-6-PDH Deficient Control (G-6-PDH CONTROL D), Catalogue Number PD2617 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Glucose-6-Phosphate Dehydrogenase in erythrocytes
FDA Recall Z-1254-2019 · Posted None
Advertisement
Recalling FirmRandox Laboratories Ltd.
Recall StatusTerminated
Product CodeJBL
Date Initiated2019-04-03
Date Terminated2019-10-09
Product Quantity14 kits
Reason for Recall
The target values and ranges in these lots are incorrect.
Firm Action
Randox sent an Urgent Medical Device Correction notification letters dated 4/2/19 to affected customers. The letter identified the affected product, problem and action to be taken:
" Discontinue use of and quarantine any of the above devices immediately.
" Discuss the contents of this notice with your Medical Director.
" Review results generated with the affected batches in line with the clinical profile of
the patient.
" Inform all relevant staff members. If you have supplied or transferred any potentially
affected product to another facility or organization, let that facility know of the recall
immediately by providing a copy of this FSN and response form.
" Complete and return the response form 12187-QA to [email protected]
within five working days.
For questions contact Randox Technical Services.
Root Cause
Nonconforming Material/Component
Distribution Pattern
The products were distributed to the following US states: IL, OH, NC, PA, WI, WV and Puerto Rico.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.