HemoSense INRatio2 Prothrombin Time Monitor, manufactured by HemoSense Incorporated. The INRatio2 PT Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood.
FDA Recall Z-1256-2009 · Posted 2009-05-22
Advertisement
Recalling FirmHemoSense Inc
Recall StatusTerminated
Product CodeGJS
Date Initiated2008-12-23
Date Terminated2011-03-25
Product Quantity921 units
Reason for Recall
INRatio Monitors that were manufactured with a specific EEPROM memory chip fail during use. The monitors report an error code and do not report a test result.
Firm Action
Letters for each product titled "Urgent: Medical Device Recall Notice" dated December 19, 2008 were issued explaining the problem and instructing users to discontinue use, complete the provided reply form and return the device by mail to Inverness Medical. Distributors of the violative device were instructed to contact their customers and provide the same recall information.
Further questions should be addressed to HemoSense Technical Service at 1-877-441-7440.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution in the US, Germany, Spain and Australia.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.