Toshiba TSX-301A AQUILION ONE (MPDCT0297EA) - Whole Body X-Ray CT Scanner
FDA Recall Z-1257-2011 · Posted 2011-02-15
Recalling FirmToshiba American Medical Systems Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2010-09-07
Date Terminated2012-04-20
Product Quantity44 units distributed between 09/2007 - 07/2010
Reason for Recall
Incorrect Dose Calculation. The software that performs the pre-scan dose calculation uses an incorrect parameter for an exam Plan using a tube voltage of 80kV or 100kV. This results in a pre-scan expected dose that is approximately 10% lower than what the displayed value should be when the scanner executes the exam Plan.
Firm Action
Toshiba America Medical Systems (TAMS) issued notification to consignees 9/7/2010, via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification.
Root Cause
Software design
Distribution Pattern
Nationwide distribution.