Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.

FDA Recall Z-1259-2018 · Posted None

Recalling FirmEncore Medical, Lp
Recall StatusCompleted
Product CodeKWS
Date Initiated2018-01-29
Product Quantity92 units

Reason for Recall

It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants

Firm Action

DJO Global sent an Urgent Field Safety Notice letter dated January 29, 2018, to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call (512) 834-6255

Root Cause

Device Design

Distribution Pattern

US Nationwide Distribution