Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device Product Code STRAP25. Intended for fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures, such as hernia repairs.

FDA Recall Z-1260-2011 · Posted 2011-02-18

Recalling FirmEthicon, Inc.
Recall StatusTerminated
Product CodeGDW
Date Initiated2010-12-29
Date Terminated2012-11-07
Product Quantity4,110

Reason for Recall

The Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device may have the potential for the sterile barrier to be compromised due to a defect in the packaging.

Firm Action

Ethicon sent Urgent Voluntary Product Recall letters dated, December 28, 2010, to their customers on December 29, 2010 via UPS. The letter identified the affected product and described the reason for recall. Customers are to discontinue use and sampling immediately. They are to return the product using the enclosed prepaid shipping label along with the enclosed business reply card regardless of whether they have any affected product or not. Customers should contact 1-877-551-7147 or speak to their manager if they have questions regarding this recall.

Root Cause

Pending

Distribution Pattern

Worldwide Distribution -- USA and Canada.