Dri-Tek BG-OX Cartridge, Model #D7120, Cat. #946-003 Cartridges used in the NPT7 Analyzer, an in vitro diagnostic device.
FDA Recall Z-1261-06 · Posted 2006-07-27
Advertisement
Recalling FirmRadiometer America Inc
Recall StatusTerminated
Product CodeCHL
Date Initiated2006-05-01
Date Terminated2012-02-24
Product Quantity185
Reason for Recall
Air bubbles may be trapped on or close to the O2 sensor in the pO2 chamber of the cuvettes in cartridges used for blood gas analysis on NPT7 analyzer. This produces too high pO2 readings without a question mark.
Firm Action
The recalling firm initiated this recall with a letter to send to their customers, via Registered Mail, on 5/1/2006. The letter requests that the customers identify and discard the potentially defective cartridges from Lot #R0299. The letter is accompanied by a recall response page to be completed by the customer and faxed or mailed back to the Radiometer so that the customer can receive an equal number of replacement cartridges free-of-charge.
Root Cause
Other
Distribution Pattern
Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.