Terminated

Philips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of Doppler velocity results values are used to assist in diagnosing the existence, location and severity of cardiac or vasculare stenosis.

FDA Recall Z-1264-2014 · Posted 2014-03-20

Recall StatusTerminated
Product CodeIYN
Date Initiated2014-02-28
Date Terminated2016-12-12
Product Quantity1267 total :670 units in US and 1267 units Internationally (which includes 11 units in Canada)

Reason for Recall

A problem has been detected with Philips Model EPIQ 5 and 7 Ultrasound System. The system will report incorrect Doppler velocities if you use High Pulse Repetition Frequency (HPRF) with multiple sample volume gates. While in High Pulse Repetition Frequency (HPRF) in Display Zoom Velocity Scale using manual spectral Doppler trace, the velocities in the velocity results table may be overstated b

Firm Action

The firm, Philips Healthcare Ultrasound, sent an "Medical Device Correction" letters dated 2014 Feb 28 via Certified Mail via the USPS by March 5, 2014. The letters described the product, problem and actions to be taken. The customers were instructed to not use HPRF in PW Doppler mode, to avoid the noted issues. Philips Field Service Engineer will be contacting them to schedule software upgrade to the systems free of charge. If you need any further information or support concerning this issue, please contact the Quality Specialist II at 978-659-4519 or your local Philips representative.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NM, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, and WV; and countries of: Australia, Austria, Belgium, Canada, Chili, Ecuador, Egypt, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Korea Rep, Malaysia, Netherlands, New Zealand, Norway, Peru, Philippines, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →