Covidien Endo GIA Ultra Universal Stapler 12 mm-intended for resection, transection in general surgery Item Code: EGIAUSTND
FDA Recall Z-1264-2020 · Posted None
Advertisement
Recalling FirmCOVIDIEN LLC
Recall StatusTerminated
Product CodeGDW
Date Initiated2020-01-17
Date Terminated2022-07-29
Product Quantity2,180 units
Reason for Recall
Potential for the sterile barrier to be compromised. Products with a compromised sterile barrier may result in a potentially increased risk for infection
Firm Action
Medtronic notified consignees by letter via Federal Express and/or certified mail beginning January 17, 2020. The letter informs of the reason for recall, health risk and action to take: Immediately quarantine and discontinue use of the affected itemcode and lot; " Return affected product; Forward the information to anyone who they have distributed. Complete the Recalled Product Return Form even if you do not have inventory.
Questions: Email Medtronic Post Market Vigilance at: [email protected].
Root Cause
Package design/selection
Distribution Pattern
US Nationwide, Israel, and Japan
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement