Impactis NaviKit-stereotaxic accessory for Computed Tomography systems Reference # I10100
FDA Recall Z-1264-2021 · Posted None
Advertisement
Recalling FirmImactis
Recall StatusTerminated
Product CodeJAK
Date Initiated2021-02-11
Date Terminated2023-10-05
Product Quantity105 kits
Reason for Recall
Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory
Firm Action
Imactis notified users by visit on 2/11/221 and issued Urgent Medical Device Recall on 2/25/21 via email to consignees. Letter states reason for recall, health risk and action to take: Customers are requested to quarantine any product from the affected NaviKit" lot numbers and return them to Imactis
Affected NaviKit lot numbers will be quarantined and removed from the market.
New compliant product will be sent to customers to replace the removed lot numbers.
As the defect is visually detectable, product will be inspected internally according to an approved procedure and any defective needle holders will be sorted and replaced with compliant ones. Refurbished NaviKit" product will not be sold to customers.
Your IMACTIS representative for questions:
Robert Hornak (VP Sales, North America) - 205-542-1486 (Cell) - 205-542-1486 (cell) Monday through Friday, 8:00 AM to 6:00 PM, Eastern Time
Root Cause
Process control
Distribution Pattern
NJ, NY, WI
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.