Terumo Advanced Perfusion System 4 inch diameter Roller Pump; Model 801040.

FDA Recall Z-1266-06 · Posted 2006-07-27

Recall StatusTerminated
Product CodeDTQ
Date Initiated2005-09-01
Date Terminated2008-06-11
Product Quantity1898 combined

Reason for Recall

The pump may stop during the autodose delivery without completion of the dose delivery to the patient.

Firm Action

Service technicians were notified on 9/30/05 to upgrade the software on these units by March 31, 2006 as follows: pump software to version 1.20, power manager software to version 1.20 and monitor software to version 2.2.1. U.S. software has all been upgraded and international upgrades are ongoing as of 7/7/06.

Root Cause

Other

Distribution Pattern

Nationwide, and worldwide, including Australia, Belgium, Canada, China, Dominican Republic, Egypt, Georgia, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand, Turkey and United Arab Emirates.