TLC-2000 Therapeutic Medical Laser System
FDA Recall Z-1266-2020 · Posted 2020-02-13
Recalling FirmTheralase Inc.
Recall StatusTerminated
Product CodeNHN
Date Initiated2019-09-26
Date Terminated2020-12-03
Product Quantity18 units
Reason for Recall
Laser Probe (Model Number: TLC-2001) Electromagnetic Interference ( EMI ) in excess of IEC-60601-1-2 (4th edition) standard
Firm Action
Email was sent out to healthcare practitioners regarding the recall on September 26, 2019.
Healthcare practitioners are instructed to courier their TLC-2000 to Theralase. The TLC-2000 will be updated to reduce EMI to meet IEC 60601-1-2 standard (4th edition).
Root Cause
Process control
Distribution Pattern
Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania