Terminated

EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.

FDA Recall Z-1267-2018 · Posted None

Recalling FirmMedica Corporation
Recall StatusTerminated
Product CodeJIT
Date Initiated2017-04-11
Date Terminated2019-08-20

Reason for Recall

Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.

Firm Action

Consignees were notified of the recall via email bulletin dated 4/11/17. The bulletin informs customers of the issues and explains the actions to be taken.

Root Cause

Process design

Distribution Pattern

USA: MI FL CA TX India, Russia, Puerto Rico, N. Mariana Island/Palau.