Terminated

GammaWin software, part number GM11019110, versions up to and including 1.62, for the MammoSource radionuclide applicator system, for radiation therapy.

FDA Recall Z-1270-06 · Posted 2006-07-27

Recall StatusTerminated
Product CodeJAQ
Date Initiated2006-05-01
Date Terminated2007-05-18
Product Quantity254 total units

Reason for Recall

Medical device software for brachytherapy may cause erroneous data to be recorded and affect patient radiation treatments. Erroneous data may be listed on the treatment history report, and the default step size may lead to a misadministration if treatment data is entered manually.

Firm Action

The recalling firm notified end users by letter on 05/01/06 and advised users of possible communication error between afterloader devices and software. The notification also advised that the control program for the devices has the potential to lead to a misadministration of treatment in cases where data is manually entered by the user. The firm additionally advised that user may continue to use the afterloader devices if following the listed cautionary statements. A software upgrade is planned to be implemented at the user site''s next regularly scheduled radiation source exchange. A response form is enclosed with the notification and users are requested to return the customer receipt verification via fax.

Root Cause

Other

Distribution Pattern

Nationwide and worldwide.