Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000C Catheterization Table CAT-880B Canon CATHETERIZATION TABLE, MODEL CAT-880B INPUT 220V - Product Usage: indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.

FDA Recall Z-1270-2021 · Posted None

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2020-12-08
Product QuantityTotal of all Systems = 145 systems (U.S.)

Reason for Recall

System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.

Firm Action

On 01/12/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter via USPS return receipt mail to customers informing them that the catheterization tables may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose. Additionally, the manual operation of the tabletop longitudinal or lateral movement will become heavy or hard to move due to the loosening of the bolts. Customers are instructed/informed: a. Should the table top tilt during a procedure before the bolts are replaced and secured, to ensure the safety of the patient, please move the table top to the position of the patients foot and refrain from operating the tabletop as much as possible. Please contact your service representative. b. Share this information with all users and reviewing radiologist as well as clinical engineering, and the biomedical group at their facility. c. Complete and return the attached Response Form and fax it to the toll free number 877-349-3054, or return via email to [email protected] For any questions or assistance, contact Regulatory Affairs Manager at 800-421-1968 or the local service representative at 800-521-1968.

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of Canada, China, Japan, Korea, and Singapore.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.