Enzyme linked immunosorbent assay, for herpes simplex virus, hsv-2
FDA Recall Z-1274-2021 · Posted 2021-03-18
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc
Recall StatusTerminated
Product CodeMYF
Date Initiated2021-02-08
Date Terminated2025-02-07
Product Quantity390 kits total (Domestic 366 kits, Foreign 24 kits)
Reason for Recall
Potential false positive HSV results. Siemens Healthcare Diagnostics, Inc. initiated a correction for ADVIA Centaur4P Herpes-2 lgG (HSV2) Onboard Stability and Calibration Frequency (Interval) Failures on the ADVIA Centaur XP System and ADVIA Centaur XPT System.
Firm Action
In February 2021, Siemens will issue an Urgent Medical Device Correction notice to customers via letter stating that Siemens Healthcare Diagnostics Inc. has confirmed that the ADVIA Centaur HSV2 assay does not meet the 60 day onboard stability or the 28 day calibration frequency (interval) as stated in the ADVIA Centaur XP/XPT Herpes-2 IgG (HSV2).
Root Cause
Under Investigation by firm
Distribution Pattern
Domestic: AL, CA, FL, IL, MD, NJ, NY, TN, TX. Foreign: Nepal, Chile, Georgia, Spain.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.