Passport V Monitor, Multiparameter Patient Monitor(with Arrhythmia Detection and Alarms) Intended for intra hospital use under the direct supervision of a licensed healthcare practitioner.

FDA Recall Z-1280-2015 · Posted 2015-03-16

Recall StatusTerminated
Product CodeMHX
Date Initiated2015-01-23
Date Terminated2017-04-20
Product Quantity519 units

Reason for Recall

An issue has been identified with Passport V Monitors invasive blood pressure function (IBP) which may provide an incorrect IBP measurement

Firm Action

Mindray DS USA Inc sent Urgent: Passport V Monitor Field Corrective Action letters dated 1/23/2015 to hospital administrators. The letters informed customers that the IBP module may provide an incorrect IBP measurement. The letter also explained the adverse effects on patients. Customers are asked to either calibrate the IBP following the instructions provided; or, contact a Mindray Service Representative, who can perform the calibration, at 1-800-288-2121. In the case of the second option, it is recommended that the monitor be removed from service until the calibration is performed. If the first option is chosen, customers are to notify a Service Representative upon completion of the action. Customers can direct their questions to Diane Arpino at 201-995-8407.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution