Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.
FDA Recall Z-1281-2015 · Posted 2015-03-16
Advertisement
Recalling FirmSynthes, Inc.
Recall StatusTerminated
Product CodeFZX
Date Initiated2015-02-17
Date Terminated2016-08-26
Product Quantity183
Reason for Recall
For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
Firm Action
An Urgent Notice: Medical Device Recall, dated February 17, 2015, was sent to all direct consignees and sales consultants instructing them to take the following actions:
Take note of the quantity of affected parts in your possession, and then follow ONE of the options for replacement of the affected product:
Option 1:
Call DePuy Synthes Repair Department at 1-800-288-6698 to schedule a repair for the affected products.
Complete the Verification Section following the instructions provided and send a copy of the completed Verification Section to DePuy Synthes by: Fax: 866-723-2823 or Scan/e-mail: [email protected]
Option 2:
Call DePuy Synthes Repair Department at 1-800-288-6698 to order repair part. The replacement part will be sent directly to your facility.
Questions should be directed to 610-719-5450 or a local DePuy Synthes sales consultant.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.