BVI Wet -Field Eraser, 18GA Blunt Tip, 45Deg.-Intended to remove tissue and control bleeding by the use of high frequency electrical current Part Number: 221251

FDA Recall Z-1281-2022 · Posted None

Recall StatusOpen, Classified
Product CodeGEI
Date Initiated2022-04-26
Product Quantity660 units

Reason for Recall

Boxes labeled as Wet-Field Eraser, 18GA Blunt Tip, 45Deg., were shipped containing incorrect product a Wet-Field Eraser, 18GA Bevel Tip, Straight, Part Number: 221250

Firm Action

BVI issued Urgent Medical Device Letter dated 4/26/22 via email stating reason for recall, health risk and action to take: 1. Immediately examine your inventory and quarantine product from these lots that are subject to recall. If you have further distributed this product, please notify your customers by initiating a sub-recall. Consider all potential users of this product in your user supply chain. You are encouraged to use a copy of this recall notification letter when contacting your customers. 2. Complete the enclosed Business Response Form (BRF) and indicate whether you have affected product to return. 3. Please enter the following information on the enclosed BRF: part number(s), lot number(s) and quantity(ies) of device(s) that are being returned. 4. Return ALL quarantined product from the affected lot to our company utilizing the Return Merchandise Authorization (RMA) sticker label that is included with this letter. You may use the prepaid BVI FedEx number for your return. If you have any specific product replacement questions you may contact our Customer Service Team at: [email protected] or call 866-906-8080 from 8 A.M. to 6 P.M. EST (Monday-Friday)

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution in the states of AZ, CA, CT, DC, FL, GA, IN, MA, MN, NV, SC, TX, WA.