Terminated

In2Bones Shannon Percutaneous Burr - Product Usage: indicated for forefoot mini-invasive and percutaneous surgery.

FDA Recall Z-1283-2021 · Posted None

Recalling FirmIn2Bones, SAS
Recall StatusTerminated
Product CodeGFF
Date Initiated2021-02-09
Date Terminated2024-07-11
Product Quantity60 units

Reason for Recall

This Field Action is being conducted following identification of manufacturing non-conformity of some batches of Shannon burrs diameter 2.0mm length 20mm, possibly causing the breakage of the burr during use.

Firm Action

A "FIELD CORRECTIVE ACTION MEDICAL DEVICE BATCH RECALL" and reply form dated 02/09/2021 were sent to customers.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the state of TN.