Neodent GM X 6mm Engaging Titanium Base, REF: 9007162

FDA Recall Z-1285-2023 · Posted 2023-03-24

Recalling FirmPreat Corp
Recall StatusOpen, Classified
Product CodeNHA
Date Initiated2023-02-21
Product Quantity26

Reason for Recall

Engaging titanium base contains a dimensional condition that allows vertical movement of the titanium base after the prosthetic screw is torqued to the recommended value, which if undetected and installed on the implant, may lead to abutment/screw fracture, which could result in the abutment/screw/crown being ingested by the patient.

Firm Action

On 2/21/23, recall notices were mailed to customers who were asked to discard or return affected devices to the recalling firm. In addition, in the event affected product was further distributed, customers were asked to inform their customers, or to provide a customer list so the recalling firm could notify these additional customers. The customer service number was provided: 800-232-7732

Root Cause

Process design

Distribution Pattern

US Nationwide distribution in the states of TX, OR, UT, NY.