Open, Classified

CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826

FDA Recall Z-1286-2023 · Posted 2024-09-30

Recalling FirmRandox Laboratories
Recall StatusOpen, Classified
Product CodeDCK
Date Initiated2023-02-01
Product Quantity2 kits

Reason for Recall

The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range

Firm Action

Randox UK (manufacturer) extended Urgent Medical Device Removal letter to the US distributor via email on 2/1/23. The distributor contacted the customer via email on 2/6/23. Letter states reason for recall, health risk and action to take: " Discontinue use of and discard any of the above immediately. Provide Randox with photographic evidence of the destruction of the kits. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to [email protected] within five working days. If you have any questions or concerns, please contact Randox Technical Services.

Root Cause

Under Investigation by firm

Distribution Pattern

PA, OH, WV