Nimbus II Infusion System; Nimbus II PainPRO IT1053; Nimbus II Flex IT1062; Nimbus II Plus IT1078; Nimbus II EpiD IT1124; Nimbus II EMS IT1063; Ambulatory Infusion Pump and components

FDA Recall Z-1286-2024 · Posted 2024-03-28

Recalling FirmInfuTronix LLC
Recall StatusOpen, Classified
Product CodeFRN
Date Initiated2024-02-21
Product Quantity52,328 total

Reason for Recall

InfuTronix is removing the Nimbus Infusion Pump Systems from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump housing. The device will not be available or supported after June 20, 2024.

Firm Action

On February 21, 2024, the firm began notifying customers via URGENT MEDICAL DEVICE REMOVAL letters. Customers were informed that the identified Nimbus Infusion Pumps are being removed from the market. The products will no longer be supported by InfuTronix for the Nimbus Infusion Pumps or related infusion sets beyond June 20, 2024. Customers of InfuTronix who have Nimbus Infusion Pumps and associated infusion sets in their possession should contact InfuTronix customer service at [email protected], where they will be instructed on the Return Material Authorization (RMA) process to be followed for returning the both the Nimbus Infusion Pumps and associated infusion sets. As the device will not be available or supported after June 20, 2024, InfuTronix encourages the healthcare provider to seek alternative methods of drug product infusion appropriate for their patient s needs, based on their medical expertise, at the earliest possible opportunity. ***Updated 4/12/24*** The firm issued updated letters on 4/9/24 to reflect that the recall had been classified Class I by the FDA. Effective immediately, Nimbus Pumps will no longer be shipped. However, administration sets will continue to be shipped through June 20, 2024.

Root Cause

Device Design

Distribution Pattern

Domestic US distribution nationwide. No international distribution.