Terminated

Signature TKA GDE/MDL Set 04-05 (Femur & Tibia) The Signature Patient-Specific Surgical Guides are intended to be used as part of the Signature Personalized Patient Care System. Lt is intended to be used as a guide during the surgical procedure for total and partial knee arthroplasty.

FDA Recall Z-1287-2015 · Posted 2015-03-17

Recalling FirmMaterialise N.V.
Recall StatusTerminated
Product CodeJWH
Date Initiated2015-02-04
Date Terminated2015-10-30
Product Quantity1 unit

Reason for Recall

Images belonging to a different patient were used for the production of the patient specific surgical guide. The incorrect guide may cause a delay in surgery to accommodate the preparation and use of traditional instrumentation.

Firm Action

Materialise N.V. sent an email on February 4, 2015, to the the affected distributor regarding the reason for the recall, affected product, and instructions for removing the affected product. For questions regarding this recall call 321-639-6611.

Root Cause

Employee error

Distribution Pattern

Worldwide Distribution to the Netherlands only.