Interventional Fluoroscopic X-Ray System - Product Usage: developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

FDA Recall Z-1288-2021 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2021-02-26
Date Terminated2021-08-11
Product Quantity29 systems

Reason for Recall

Siemens has become aware of a potential issue in which the imaging system UPS sporadically forces a shutdown of the imaging system PC without a true power supply problem.

Firm Action

On 2/26/2021, Siemens issued an Urgent Medical Device Correction notice to in inform customers of a possible hazard to patients, operators, or other persons and equipment due to a hardware error which affects ARTIS icono /pheno systems in combination with imaging system UPS (Uninterrupted Power Supply) option. This potential hardware issue has no influence on the previous treatment/diagnosis of patients. However, in case of error, planned procedures may have to be terminated and performed on an alternative x-ray system. Based on our investigation, you can continue to safely use your system.

Root Cause

Component design/selection

Distribution Pattern

US Nationwide distribution in the states of MI, KY, TX.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.