Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Corin BIOLOX DELTA MOD HEAD 32mm EX LONG +7mm 12/14 TAPER. Corin Trinity Acetabular System with Extra Long Heads/Trinity Biolox Delta Ceramic Head (Generic Name: Hip Prosthesis). The Trinity Acetabular System is a modular acetabular cup system consisting of a press fit, titanium alloy shell; acetabular liners in neutral offset, +4mm offset, +4mm oblique, neutral 4mm EPW; and ceramic and CoCrMo modular heads. The ceramic and CoCrMo modular heads are compatible with Corin titanium femoral stems. The Trinity Biolox Ceramic heads are packaged in a double peel pouch configuration.

FDA Recall Z-1290-2021 · Posted None

Recalling FirmCorin Ltd
Recall StatusTerminated
Product CodeMEH
Date Initiated2021-01-25
Date Terminated2022-11-23
Product Quantity23 units

Reason for Recall

The size indicated on the labeling on the outer packaging, on the inner packaging, and on the patient stickers inside the packaging is incorrect: the Trinity Biolox Delta Ceramix Head 32XL is labeled as 36XL and vice-versa.

Firm Action

The firm contacted their consignee by email on 01/25/2021. The distributor was instructed to identify the affected product in its inventory and return it to Corin USA in order to contain the part. A recall letter will be sent and acknowledgement of receipt will be requested.

Root Cause

Labeling mix-ups

Distribution Pattern

Worldwide distribution - US Nationwide distribution and the countries of Colombia, Poland, Germany, Australia.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑