GE Medical Systems COROMETRICS Maternal/Fetal Monitoring Systems, Models 2126is, 2128is, 2129is

FDA Recall Z-1291-03 · Posted 2003-09-23

Recall StatusTerminated
Product CodeHGM
Date Initiated2003-05-21
Date Terminated2003-10-27
Product Quantity170

Reason for Recall

The 2120 Main Board on device lacks required external safety 'watchdog' circuit.

Firm Action

The firm mailed or faxed a recall letter to consignees on May 14, 2003 followed by a phone call to each account within two weeks to arrange for a service technician visit to correct the monitors. The recall letter warned the hospitals of the problem and a work around.

Root Cause

Other

Distribution Pattern

The monitors were distributed to 8 hospitals in SC, WI, CA, MN, TX, MD and NV, as well as two hospitals in France.