DS Titanium Ligation Clip Appliers, used in laparoscopic surgery
FDA Recall Z-1291-2018 · Posted None
Advertisement
Recalling FirmAesculap Implant Systems LLC
Recall StatusTerminated
Product CodeFZP
Date Initiated2018-01-22
Date Terminated2019-09-24
Product Quantity270
Reason for Recall
Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. Updated cleaning and sterilization instructions.
Firm Action
A letter and copy of the updated IFU was sent to affected customers the week of January 15, 2018. Product does not need to be returned.
IFU is revised and put on the Aesculap Website for usage. https://www.aesculapusa.com/products/instructions-for-use.
Root Cause
Labeling design
Distribution Pattern
US and Canada distribution