Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

PCRopsis BCSNano (1mL), REF 2276001; nucleic acid extraction from a variety of biological samples

FDA Recall Z-1295-2023 · Posted 2023-03-29

Recalling FirmEntopsis , Inc.
Recall StatusOpen, Classified
Product CodeJJH
Date Initiated2023-02-21
Product Quantity3 units

Reason for Recall

Presence of misleading label statements on the product label.

Firm Action

The firm issued a MEDICAL DEVICE RECALL notice to its consignees on 02/21/22023 by email. Those without a valid email address were mailed the notice via certified mail. The notice explained the issue and risk and requested the customer/user take the following actions: "Entopsis Inc., requests that the product lots listed below - shipped/received before February 8st, 2023 be promptly discontinued from use and discarded from inventory as per your non-hazardous waste disposal plan.The labeling of PCRopsis products was revised on February 8th, 2022. Updated statements of intended use and revised instructions for use are available on www.pcropsis.com. If you are currently using or intend to use the PCRopsis product lots affected by this recall please contact us at [email protected] or 1-888-407-5070 (Monday to Friday 9.00 AM to 4.30 PM Eastern Time) to request a replacement. If you are a re-distributor of PCRopsis devices, notify your end customer with a copy of this letter. Please acknowledge the removal of affected product lots by filling out the attached acknowledgment form and returning it to us at [email protected]. Contact us at [email protected] or 1-888-407-5070 - Monday to Friday 9.00 AM to 4.30 PM US Eastern Time (GMT -5) with any questions you may have."

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide and Canada, UK, Portugal, England, Spain, England , Germany, Brazil, Denmark, KOR, South Korea, Colombia, Lithuania

Official Source

View FDA Recall →