Terminated

Cios Alpha, Model Number 10308191 Mobile X-ray system - Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.

FDA Recall Z-1298-2019 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2019-03-18
Date Terminated2020-04-16
Product Quantity257

Reason for Recall

Continuous operation at very high tube output may result in increased wear of the radiation emitting components. This may occur when tube voltages of 125 kV are used for a long period, e.g. 30 minutes or more of continuous use. As a result, risk of premature failure of the Monoblock and loss of imaging X-rays.

Firm Action

Urgent Medical Device Correction letters dated 3/18/19 were sent to customers.

Root Cause

Device Design

Distribution Pattern

US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI.