Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet Mapping Assay - PlateletMapping ADP & AA (07-014), PlateletMapping ADP (07-015), PlateletMapping AA (07-016). Issue occurs when Platelet Mapping Assay results from a TEG 5000 analyzer are viewed using TEG Manager Software (versions 1.1-4.3.1)
FDA Recall Z-1298-2022 · Posted None
Advertisement
Recalling FirmHaemonetics Corporation
Recall StatusTerminated
Product CodeJPA
Date Initiated2022-04-19
Date Terminated2024-03-27
Product Quantity111
Reason for Recall
When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
Firm Action
On April 22, 2022, the firm notified affected customers of the recall via email.
Customers were asked to complete and return the attached acknowledgement form. Upon receipt, Haemonetics will contact the customer to arrange the remote installation of the updated software, version 4.4.0, which will resolve the issue.
Root Cause
Software design
Distribution Pattern
US Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.