Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10
FDA Recall Z-1298-2026 · Posted 2026-02-05
Advertisement
Recalling FirmWilson-Cook Medical Inc.
Recall StatusOpen, Classified
Product CodeKNT
Date Initiated2025-12-15
Product Quantity60 units
Reason for Recall
Nasal feeding tube packaged without the nasal transfer tube component.
Firm Action
On January 5, 2026, Cook Endoscopy issued a Urgent Medical Device Recall Notification to affected consignees via E-Mail. Cook Endoscopy asked consignees to take the following actions:
1. Examine inventory immediately to determine if you have affected product(s). If so, quarantine any affected products that remain unused. Immediately cease all further distribution and use of the affected products.
2. Return the affected product(s) to Cook Medical with a signed copy of the Acknowledgement and Receipt Form.
3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand and email the completed form to [email protected].
4. Please share this notice with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected products have been transferred.
Root Cause
Employee error
Distribution Pattern
International distribution to the country of Australia.