Terminated

The ImmunoCard Mycoplasma EIA detects the presence of IgM to M. pneumoniae in serum - Product Usage: intended to aid in the diagnosis of recent Mycoplasma pneumoniae infection.

FDA Recall Z-1299-2021 · Posted None

Recall StatusTerminated
Product CodeLJZ
Date Initiated2021-02-08
Date Terminated2021-10-26
Product Quantity8 kits

Reason for Recall

Use of the kits after incorrect storage could lead to a potential false-negative or invalid result. In the event of a false-negative, the incorrect assumption would be made as to the cause of patient symptoms. The appropriate treatment may not be administered, and symptoms may progress. Additional evaluations will be completed and documented within the associated Health Hazard Evaluation.

Firm Action

On 02/19/2021, Meridian issued an Urgent Medical Device Correction notice to customers via letter for ImmunoCard Mycoplasma kits from a specified lot. Upon receipt at the Cardinal Health Charlotte distribution center, the kits were identified to have a transit time of 115 hours with a maximum temperature of 55 F (approximately 12 C). This is beyond our transit time requirement of 72 hours and storage at 2-8 C. Cardinal placed the kits into a quarantine location, however 8 kits were inadvertently shipped to your facility.

Root Cause

Storage

Distribution Pattern

US Nationwide distribution in the states of OH, SC.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →