RayStation 11B
FDA Recall Z-1299-2022 · Posted None
Advertisement
Recalling FirmRAYSEARCH LABORATORIES AB
Recall StatusTerminated
Product CodeMUJ
Date Initiated2022-06-10
Date Terminated2024-06-11
Product Quantity26
Reason for Recall
This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation
11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display
might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync
with the selected beam.
Firm Action
The firm notified customers of the recall via email on June 10, 2022. The notice was titled "Field Safety Notice, Medical Device Correction #99834."
Customers were given precautions and instructed on a workaround.
The issue will be resolved in the next version of RayStation, scheduled for market release in June 2022, though this may be subject to market clearance.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the states of NJ and TX.