FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
FDA Recall Z-1299-2026 · Posted 2026-02-05
Advertisement
Recalling FirmFUJIFILM Healthcare Americas Corporation
Recall StatusOpen, Classified
Product CodeKPR
Date Initiated2026-01-09
Product Quantity16 units
Reason for Recall
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Firm Action
On January 9, 2026 URGENT: MEDICAL DEVICE RECALL letters were emailed to customers. Actions to be taken:
Please pay attention to the following points until the work is completed.
" For safety reasons, please turn off the breaker of the distribution board to which the power cord of the unit is connected after turning off the power of the unit according to the operation manual.
" Do not forget to turn off the breaker of the distribution board until the replacement of the contactor is completed.
1) Please post this notification on or near the device
2) Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken.
3) Please contact Technical Assistance Center (TAC) at 1-888-FUJIMED (1-888-385-4633) to open a service ticket.
4) Please report any known adverse events to FDA via MedWatch.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.