Boston Scientific Urovac Bladder Evacuator, sterile, single use only; REF/Catalog no. 730-125 and sold individually as UPN M0067301250 and in packaged of ten as UPN M0067301251.

FDA Recall Z-1300-06 · Posted 2006-07-27

Recall StatusTerminated
Product CodeKOG
Date Initiated2006-06-30
Date Terminated2008-07-31
Product Quantity446,097 units

Reason for Recall

Sterility is compromised because the product is labeled as sterile, but the packages may have holes in them.

Firm Action

U.S. accounts were notified via recall letter dated 6/30/06 to cease use of the product and to return it to the firm. International accounts were also notified.

Root Cause

Package design/selection

Distribution Pattern

Nationwide and worldwide, including Argentina, Bangladesh, Belgium, Canada, Chile, China, Colombia, Dominican Republic, Egypt, El Salvador, Finland, France, Germany, Great Britain, Greece, Hong Kong, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Portugal, Romania, South Africa, Spain, Sri Lanka, Sweden, Tunisia, Turkey and Venezuela.