Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Confidence Plus Kit Spinal Cement System, Product Codes: 283910000 (GTIN: (01)10705034209623) and 283910000 (GTIN: (01)10705034209630) The CONFIDENCE SPINAL CEMENT SYSTEM 11cc Kit is intended for percutaneous delivery of CONFIDENCE 11cc Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancer, myeloma).

FDA Recall Z-1300-2018 · Posted None

Recalling FirmDePuy Spine, Inc.
Recall StatusTerminated
Product CodeNDN
Date Initiated2017-12-15
Date Terminated2020-04-13
Product Quantity1972

Reason for Recall

Certain lots of the spinal cement may contain incorrect amounts of contrast medium so that the product is difficult or impossible to visualize on x-ray.

Firm Action

The firm, DePuy Synthes, sent an "URGENT FIELD SAFETY NOTIFICATION-RECALL CONFIDENCE KIT SPINAL CEMENT SYSTEM" dated 12/2017 to its consignees. The notification described the product, problem and actions to be taken. Notifications were sent to consignees with the following instructions: 1. Immediately identify and quarantine all unused products listed below in a manner that ensures the affected products will not be used. 2. Review, complete, sign and return the attached Business Reply Form to your local DePuy Synthes sales organization in accordance with the directions on the form within 5 business days of receipt of this notification. 3. Return any affected product within 30 business days. A credit note will be issued for the returned items. 4. Forward this notice to anyone in your facility that needs to be informed. 5. If any impacted products have been forwarded to another facility, contact that facility and provide them with this letter. 6. Maintain awareness of this notice until all products listed below have been returned to DePuy Spine. 7. Keep a copy of this notice. If you have questions about this Field Safety Notification, please contact your DePuy Synthes Spine Sales Consultant or the Recall Coordinator via telephone: 1-508-828-6609 or via email: [email protected]).

Root Cause

Nonconforming Material/Component

Distribution Pattern

International Distribution (outside the U.S.) only: Mexico, Canada, Ireland, Great Britain, Malaysia, Germany, Austria, Colombia, Spain, Austria, Switzerland, Australia, Hungary, Croatia, and Sweden.